Troubleshooting Guide
Problem
The distiller is leaking.
Use and Care Guide
(Continued)
Possible cause
The distiller is not in a
well ventilated area.
There is excess scale
build-up in the boiling
chamber.
The boiling chamber is
not properly installed.
The O-ring is damaged
or missing from the
baffle.
The flat gasket between
the boiling chamber
spout and baffle is
missing or damaged.
Solution
Distiller must be
operated in a well
ventilated area for proper
heat dissipation.
See page 4.
Clean the boiling
chamber. See page 8.
Push the boiling
chamber into the distiller
chassis so that both the
electrical and the baffle
connection are tightly
joined. See pages
5 and 6.
If damaged, replace
O-ring. Make sure the
O-ring is properly placed
on the baffle, fitting in
the circular groove.
See page 6.
Replace gasket
10