The device is made up of a lead battery which cannot be accessed by outside. In order to replace it, consult the technical staff
authorised by the manufacturer.
USE ONLY THE RECOMMENDED BATTERIES FROM CA-MI. THE USE OF OTHER BATTERIES ARE NOT
RECOMMENDED AND INVOLVING THE CANCELLATION OF WARRANTY
In the event that the service personnel has to replace the internal battery, pay special attention to the polarity of the same
component. The + / - polarities are indicated directly on the battery.
If the overfill security system it's activated, don't proceede with the liquid aspiration. If the overfill security system doesn't work
there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter who avoid the liquid
penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the device, in this case return the
device to CA-MI technical service.
CA-MI Srl will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information
that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
CA-MI TECHNICAL SERVICE. CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.
The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or
storage.
The working position must be such as to allow one to reach the control panel and to have a good view of the empty
indicator, the jar and the antibacterial filter.
It is recommended not to keep the device in your hands and / or to avoid prolonged contact with the body of apparatus.
WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full volume of use of the jar and
better efficiency of the overflow device.
Place the unit on a flat, horizontal surface
Connect one end of the short silicon tube, with antibacterial filter, to the suction connector on the lid of the jar.
The other tube already connected to the filter has to be connected to the "VACUUM" jar outlet, where has been fixed the red
float (security float). When the 90% of the volume of the jar is reached there is the activation of the security float (the float
close the aspiration connector on the jar) to avoid liquid penetration inside the device.
WARNING: Ensure that the FLUID SIDE or IN marker on the filter is on the side facing the collection jar lid and fitted into the
"VACUUM". A wrong connection causes immediate destruction in case of contact with sucked liquids.
Filter assembling
Suction Pump Inlet
Connect the long silicone tube to the "PATIENT" jar outlet
Connect the other end of the long silicon tube to the probe plastic connector, then connect the suction probe to it.
Connect the power cord to the device then connect the plug to the electrical mains supply.
Push switch on position I to start suction.
Unscrew the lid of the jar and fill the jar 1/3 full or ordinary water (this assists the unit to reach peak vacuum performance
and makes clean-up easier) then re-screw the lid on the jar correctly.
During operation the jar has to be in vertical position to avoid overflow valve to cut off aspiration. Should this happen,
switch off the device and disconnect the tube from the jar cover (from "VACUUM" outlet).
Once finished push switch on O position and unplug.
Remove the accessories and clean.
At the end of each use, place the device in its box away from dust.
INSTRUCTION FOR USE
FLOW DIRECTION
IN
20
IN / Fluid Inside
GB