Mains Operating Voltage
Check that the instrument operating voltage marked on the rear panel is suitable for the local
supply. Should it be necessary to change the operating voltage, proceed as follows:
1.
Ensure that the instrument is disconnected from the AC supply.
2.
Remove the screws holding the case upper and handle and lift off the case upper.
3.
Change the transformer connections following the diagram below:
4.
Re-assemble in the reverse order.
5.
To comply with safety standard requirements the operating voltage marked on the rear panel
must be changed to clearly show the new voltage setting.
Fuse
The AC fuse is located in the fuse drawer in the lower part of the IEC inlet connector. To change
the fuse remove the line cord and open the fuse drawer with a suitable tool.
The correct fuse type is 20 x 5mm 250V HBC time-lag with the following rating:
Make sure that only fuses with the required current rating and of the specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuseholders are prohibited.
Mains Lead
Connect the instrument to the AC supply using the mains lead provided. Should a mains plug be
required for a different mains outlet socket, a suitably rated and approved mains lead set should
be used which is fitted with the required wall plug and an IEC60320 C13 connector for the
instrument end. To determine the minimum current rating of the lead-set for the intended AC
supply, refer to the power rating information on the equipment or in the Specification.
Any interruption of the mains earth conductor inside or outside the instrument will make the
instrument dangerous. Intentional interruption is prohibited.
230V Operation
BLUE
BROWN
230V operation :
115V operation:
WARNING! THIS INSTRUMENT MUST BE EARTHED.
115V Operation
BLUE
1.6A (T)
3.15A (T)
Installation
BROWN
5