Aim TTi EL302P Manuel D'instructions page 12

Publicité

Les langues disponibles
  • FR

Les langues disponibles

I?
VO?
IO?
OUT?
M?
ERR?
Miscellaneous Commands
*RST
*IDN?
Returns the output current limit in Amps in <nr2> numeric format.
The syntax of the response is
Example:
If the current limit is 1.00 Amps the response to the
command I? will be I 1.00<rmt>.
Reads and returns the actual output voltage in Volts in <nr2>
numeric format. The syntax of the response is
Example:
If the actual output voltage is 12.55 Volts the
response to the command VO? will be V12.55<rmt>.
Reads and returns the actual output current in Amps in the <nr2>
numeric format. The syntax of the response is
Example:
If the output current is 0.93 Amps the response to
the command IO? will be A0.93<rmt>.
Returns status of the output as
or
Returns output mode status as
or
Returns the value in the error register in the form ERR <nr1><rmt>
0 = no errors.
1 = command not recognised.
2 = command value outside instrument limits.
Resets the instrument to the default power-up settings
(1.00V, 1.00A, output off).
Returns the instrument identification. The exact response is
determined by the instrument configuration and is of the form
<NAME>,<model>, 0, <version><rmt> where <NAME> is the
manufacturer's name, <model> defines the type of instrument and
<version> is the revision level of the software installed.
I<nr2><rmt>
<nr2>V<rmt>
<nr2>A<rmt>
OUT ON<rmt>
OUT OFF<rmt>
M CV<rmt>
M CC<rmt>