Troubleshooting
5
Troubleshooting
Error
Sealing frame leaking
Nose piece guide leaking
Exhalation valve leaking
Facepiece leaking
Plug-in connector for
residual pressure alarm
leaking
Cause
Sealing frame is partly on
the head cover
Sealing frame has buckled Slide your finger along the underneath of the sealing
Head harness loose
Sealing frame is faulty
O-rings faulty
Seat or valve disc dirty or
damaged
O-ring connector faulty
Facepiece not engaged
correctly
Connector contaminated
Connector on the side of
the sealing frame leaking
Screw connection of the
connector is loose
Remedy
Check with the fingers and, if necessary, pull parts of the
head cover forwards under the sealing frame.
frame.
Re-tighten individual straps. For further information, see
the following chapter: "Placing the full face mask on the
head", page 19.
Have the sealing frame replaced by authorised specialist
personnel.
Have them checked by specialist personnel and repaired
if necessary
Visually inspect exhalation valve and replace or clean as
necessary. For further information, see the following chap-
ter: "Maintenance", page 23.
Replace the O-ring
Check for a correct fit; if necessary, re-insert the connec-
tor until the spring audibly engages in the plug-in connec-
tor
Cleaning
Tighten the nut inside
Grasp the connector on the outside of the visor and
tighten it in a clockwise direction (only gently by hand!)