Operation
5.2 Sensor setup menu (continued)
Change measurement
range
Update filtration probe
Change electrode type
Fieldbus
Option
Validation
5.2.1 System setup menu
For more information on System Setup (current outputs, relays and network interfaces),
refer to the sc1500 user manual.
5.3 Calibration process
Note: Make sure that all solutions are available to avoid incorrect measurements.
1. To start an automatically calibration select
OR
1. To start a calibration manually select CALIBRATION>CALIBRATION.
Depending on the number of standards requiring calibration, a cycle may take up to
40 minutes. After a successful calibration, the instrument automatically returns to the
measurements.
Note: When the instrument detects and displays a warning, the measurement process will continue.
Refer to
Note: When the instrument detects and displays an error, the instrument stops measuring. Refer to
section 7.2.2 on page 57
5.4 Cleaning process
For information about the cleaning interval refer to
Note: Make sure that the cleaning solution is available that the instrument can work properly.
1. To configure an automatical cleaning interval select
10
All manuals and user guides at all-guides.com
Software changes to another measuring range: ATTENTION, the appropriate standards must
be used!
Enables the filtration probe software to be updated.
Not used at the moment
ENABLED/DISABLED: Enable external control of instrument by Fieldbus. NOTE: When
instrument is put into SERVICE MODE by menu, the Fieldbus control is temporarily disabled.
Sets the instrument to filter-probe/ 1channel/ 2channel mode. Switching the options needs
modification of hardware!
Menu based process to measure external samples. When "Modification required" is
displayed: Disconnect sample tubing from overflow vessel, plug overflow vessel and put
sample tubing into external sample. After process: Unplug overflow vessel and reconnect
sample tube.
CALIBRATION>CALIBRATION>AUTOMATIC CALIBRATION>INTERVAL.
Note: Press START to confirm and start the calibration process.
After the next measurement and the 5 minutes waiting time, the calibration will
automatically start and proceed with all necessary standards.
section 7.2.3 on page 60
to troubleshoot the error.
SETTINGS>CLEANING>INTERVAL.
to troubleshoot the warning.
Table 4 on page
46.