GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS
The digit Power
Dispenser is intended for use in the electromagnetic environment specified below. The customer or the end user of the digit Power
®
Dispenser should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people
50°C
122°F
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the digit Power
Caution, consult
is used exceeds the applicable RF compliance level above, the digit Power
For Dental Use Only
instructions for use
performance is observed, additional measures may be necessary, such as reorienting or relocating the digit Power
range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
RECOMMENDED SEPARATION DISTANCE BETWEEN PORTABLE & MOBILE RF COMMUNICATIONS EQUIPMENT & THE MODEL
The medical electric equipment (digit Power
controlled.
The customer or the user of the digit Power
portable and mobile RF communications equipment (transmitters) and the medical electric equipment (digit Power
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Compliance
Electromagnetic environment - guidance
level
3 V
Portable and mobile RF communications equipment should be used no closer to any
3 V/m
part of the Medical Electric Equipment (digit Power
the recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = [3.5/V1] √P
d = [3.5/E1] √P 80 MHz to 800 MHz
d = [7/E1]√P 800 MHz to 2.5 GHz
Where
to the transmitter manufacturer and
(m) Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey, a should be less than the compliance level in each frequency range.b Interference
may occur in the vicinity of equipment marked with the following symbol:
5V-1A
5%
www.dentsplysirona.com
Keep Dry
www.dentsply.eu/ifu
Fragile
Dispenser) is intended for use in an electromagnetic environment in which radiated RF disturbances are
®
Dispenser can help prevent electromagnetic interference by maintaining a minimum distance between
®
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
[ ]
3.5
√P
d=
E
1
0.12
0.37
1.17
3.69
11.67
P
is the maximum output power rating of the transmitter in watts (W) according
106
kPa
95%
%
80
kPa
Atmospheric
Relative Humidity
Dispenser should be observed to verify normal operation. If abnormal
®
Pressure Limitation
13Va
m
80 MHz to 800 MHz
[ ]
3.5
√P
d=
E
1
0.12
0.37
1.17
3.69
11.67
Dispenser), including cables, than
®
d
is the recommended separation distance in metres
Dispenser. b Over the frequency
®
Dispenser) as recommended
®
800 MHz to 2.5 GHz
[ ]
7
√P
d=
E
1
0.23
0.74
2.33
7.38
23.33
®