Télécharger Imprimer la page

Arthrex AR-1650-02 Manuel D'instructions page 7

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Instructions for Use
Table 1: Validated Washer Parameters for the AR-1650-02 & AR-1650-03
Phase
Pre-Wash 1
Wash 1
Rinse 1
Rinse 2
Thermal
Disinfection
Drying
Steam Sterilization for AR-1650-02 & AR-1650-03
Sterilizer Type:
Preconditioning Pulses:
Minimum Temperature:
Full Cycle Time:
Minimum Dry Time:
Cool-Down Time:
Sterilizer Type:
Minimum Temperature:
Full Cycle Time:
Minimum Dry Time:
Cool-Down Time:
©2019 Allen Medical Systems, Inc. All Rights Reserved
Document Number: D-720771
Version B
Recirculation time
(minutes)
3:00
10:00
3:00
3:00
5:00
6:00
Prevacuum
4
132°C
4 minutes
30 minutes
30 Minutes outside of chamber on a wire rack
Gravity
132°C
15 Minutes
50 Minutes
30 minutes outside of chamber on a wire rack
Temperature
Detergent type and
Cold tap water
65.5°C
50°C
50°C
90°C
115°C
Ref Blank Template: 80026387 Ver. B
Concentration
N/A
Neodisher®
Mediclean Forte
¼ oz/gallon
N/A
N/A
N/A
N/A
Issue Date: 21 MAY 2020