Arthrex AR-1650-02 Manuel D'instructions page 6

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Instructions for Use
Cleaning and Sterilization Directions for Use for Reusable Connection Devices for Use
with AR-1650 Shoulder Suspension System
AR-1650-02
Arm Sleeve Connector
The Cleaning and Sterilization directions apply to the S3 Arm Sleeve Connector (AR-1650-
02) and the S3 Lateral Traction Sling Connector (AR-1650-03). For further directions for use
regarding the Arthrex
720764).
WARNING: Failure to properly sterilize device could cause exposure to contaminant.
Automatic Cleaning Procedure
AR-1650-02 & AR-1650-03
1. Prepare the detergent Neodisher® Mediclean Forte at ¼ oz/per gallon of
lukewarm tap water in an ultrasonic bath.
2. Fully immerse the devices in the prepared detergent and sonicate for 10 minutes.
3. Remove the devices from the detergent and transfer them into the washer for
processing.
4. The washer parameters that have been validated are shown below in Table 1.
The motor speed should be set to high.
©2019 Allen Medical Systems, Inc. All Rights Reserved
Document Number: D-720771
Version B
AR-1650-03
Pictures are for visual reference only
Shoulder Suspension System please refer to the S3 System IFU (D-
®
Lateral Traction Sleeve Connector
Ref Blank Template: 80026387 Ver. B
Issue Date: 21 MAY 2020
Page 6