Setting 3D calibration without using 3D glasses:
⇒ Leave the EASY view 3D in the previously described positi-
on.
EN
⇒ Switch the head off at the switch (7).
The following procedure can take up to 20 seconds:
⇒ While holding down the right mouse button and the mouse
wheel, switch on the head again, keep holding both the right
mouse button and the mouse wheel down and simultane-
ously move the mouse until the service menu appears in the
top right edge of the monitor.
Service menu:
• Firmware version
• Offset calibration
⇒ Using the mouse wheel, select "Offset calibration" and open
with the left mouse button.
Setting x axis:
⇒ Using the mouse wheel, select the x axis and open with the left mouse button.
♦ The offset value of the x axis will be displayed in a monitor window in the top right-hand corner.
⇒ Use the mouse wheel to change the value.
♦ The vertical bar of the cross is visible in duplicate.
⇒ Use the mouse wheel to set the value so that the two bars
are placed precisely over one another.
⇒ Confirm the setting using the left mouse button.
Setting the y axis:
⇒ Using the mouse wheel, select the y axis and open with the left mouse button.
♦ The offset value of the y axis is displayed in a monitor window in the top right-hand corner.
⇒ Use the mouse wheel to change the value.
♦ The horizontal bar of the cross is visible in duplicate.
⇒ Use the mouse wheel to set the value so that the two bars
are placed precisely over one another.
⇒ Confirm the setting using the left mouse button.
Leaving the 3D calibration:
⇒ Click the right mouse button.
♦ Jump back to the service menu.
⇒ Click the right mouse button.
♦ Leaving the service menu.
⇒ Switch the head off and then back on again (7, Fig.).
Check the 3D calibration using 3D glasses:
⇒ Switch to 3D display.
♦ If a clear 3D image is detected, then the 3D calibration is correct.
♦ If no clear 3D image is detected, repeat the 3D calibration.