Recommended separation distances between portable and mobile RF communications equipment and the AC-120MD
The AC-120MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the AC-120MD can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the AC-120MD as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output power of transmitter
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Caution
When you dispose of the unit or accessories, you must
obey the laws in the relative area or country and the
regulations in the relative hospital regarding
environmental pollution.
Warning on power connection
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
Appliance Connector / Plug with earthing-contacts
that conforms to the safety regulations of each
country if applicable.
2. Use the Power Cord (3-core mains lead) / Appliance
Connector / Plug conforming to the proper ratings
(Voltage, Ampere).
If you have questions on the use of the above Power
Cord / Appliance Connector / Plug, please consult a
qualified service personnel.
For the customers in U.S.A. and Canada
Warning on power connection for medical use
Please use the following power supply cord. With
connectors (plug or female) and cord types other than
those indicated in this table, use the power supply cord
that is approved for use in your area.
6
GB
W
0.01
0.1
1
10
100
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
Plug Type
Cord Type
Minimum Rating for Plug and
Appliance Couplers
Safety Approval
*Note: Grounding reliability can only be achieved when the equipment is
connected to an equivalent receptacle marked 'Hospital Only' or 'Hospital
Grade' .
Warning
The apparatus shall not be exposed to dripping or
splashing. No objects filled with liquids, such as vases,
shall be placed on the apparatus.
Do not install the appliance in a confined space, such as
book case or built-in cabinet.
Інформація для споживачів в Україні.
Обладнання відповідає вимогам:
– Технічного регламенту обмеження використання
деяких небезпечних речовин в електричному та
електронному обладнанні (постанова КМУ від 03/
12/2008 № 1057).
m
800 MHz to 2.5 GHz
d = 1.2
d = 2.3
0.12
0.38
1.2
3.8
12
United States and Canada
HOSPITAL GRADE*
Min. Type SJT
Min. 18 AWG
10A/125V
UL Listed and CSA
0.23
0.73
2.3
7.3
23