  • FR

Les langues disponibles

  • FRANÇAIS, page 27
4. Select an option.
Option
Export Data
Export Settings Saves a copy of the user settings to the USB flash drive. Select a filename
Import Settings Shows the configuration files that were saved to the USB flash drive. Selects
USB Drive Info
4.6 Use FSDATA Desktop (optional)
Use
DATA Desktop to look at sampler data or make reports. Before this task, become familiar with
FS
the
DATA Desktop menus and navigation. Refer to the
FS
Use a PC with
DATA Desktop to look at sampler data on a USB flash drive. As an alternative, use
FS
a USB A to A cable to connect the sampler to a PC with

4.7 Manual operation

Use manual operation to collect a grab sample, move the distributor arm or operate the pump. For
additional information, refer to the expanded version of this manual on the manufacturer's website.
Note: Temporarily stop the sampling program before manual operation. Push RUN/HALT, then select Halt
Program.
Section 5 Troubleshooting
Problem
The red indicator light
flashes.
"---" shows on the
Status screen.
Description
Saves a copy of the user settings, log files, sample history and manufacturer
settings to the USB flash drive. The file size of the data saved to the USB
flash drive shows on the display.
Note: Use
DATA Desktop to look at the files.
FS
for the user settings. Options: Settings 1 to 10.
a configuration file and saves it to the controller.
Shows the total, used and free (available) memory on the USB flash drive.
Possible cause
One or more
measurements that the
sampler is configured to
use (e.g., pH and flow) is
not available because the
sensor(s) is not
connected to the
sampler.
The measurement is not
available or has not been
recorded yet.
DATA Desktop documentation.
FS
DATA Desktop.
FS
Solution
Connect the missing sensor to the sampler or
make sure that the sampler is not configured:
• To record sensor measurements to the data
log that are not available.
• With alarms for sensor measurements that
are not available.
The measured values shown are the last
recorded value. Measurement data is recorded
to the data log in accordance with the selected
logging interval.
Make sure that the measurement is supplied to
the sampler.
English 25