USE OF NEBULIZATION MANUAL CONTROL
To achieve continuous nebulisation action you should not use the manual nebulisation
control (10), especially in the case of children or persons with reduced physical, sensory,
or mental capabilities. The manual nebulisation control is useful for limiting dispersion of
the medication in the surrounding environment.
To start nebulizing close with a
finger the hole of the nebulizer
manual control (10) and breathe
in gently; we recommend to
hold your breath for a moment
so that the inhaled aerosol drop-
lets can be deposited,
CLEANING SANITISATION DISINFECTION STERILISATION
Switch off the device before any cleaning procedure and unplug the power cable from
the socket.
DEVICE AND TUBING EXTERIOR (B)
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any
sort).
ACCESSORIES
Open the nebuliser by turning the upper part (13.3) anticlockwise, remove the nozzle
(13.2) from the upper part (13.3). Disassembly by pressing as shown by the 2 arrows in the
"Connection diagram" in point 13.2.
Then proceed according to the following instructions.
CLEANING AT HOME - SANITISATION AND DISINFECTION
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one of
the methods described below.
method A: Sanitise accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4
table hot water (approximately 40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4 in the dish-
washer with a hot cycle.
method C: Sanitise accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4 by immers-
ing them in a solution of 50% water and 50% white vinegar, then rinse thoroughly under
potable hot water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After having sanitised the accessories, shake them vigorously and place them on a paper
towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the
methods described below.
method A: Accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4 can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlo-
rite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
meanwhile, to disable nebuliz-
ing, remove your finger from the
hole of the nebulizer manual
control to avoid the waste of
drug, optimizing its acquisition.
Then exhale slowly.