CP 50 and CP 50 Plus 12-lead resting electrocardiographs — software update 01.30.04
Overview
This document describes the steps required to correctly update CP 50 devices to the latest software
version using the UpdateUtility. It is important that you follow the following flow chart to identify the
correct section to use when performing your update. The flow chart will determine the correct software
update kit to use.
See "Updating your CP 50 software to
version 1.30.04 from Operating mode" on
page 3
Start
Power on the device
Write down your software version
Make note of the
mode that the
device enters
See "Updating your CP 50 software to
version 01.30.04 from ATE mode" on
page 4
• For software versions 01.20.01 and earlier,
use software update kit 411053
• For software versions 01.20.02 and later,
use software update kit 411054
See "Updating your CP 50 software to version
01.30.04 from Maintenance mode" on page 6