Temperature measurement
WARNING
temperature probe to live circuits.
Insert the thermocouple probe adapter (P) with the - side
going into the COM input terminal (I) and the + side going
into the V input terminal (M).
Connect the thermocouple probe (Q) to the thermocouple
probe adapter (P). The polarity marking of the probe (Q)
must correspond to the marking of the adapter (P). The
probe pins are made in varying sizes to avoid a wrong
connection.
Set the function switch (K) to the TEMP/°C/°F position.
Press the MODE/VFD button (H) to change the temperature
unit (°C/°F) shown on the display (F).
Touch the thermocouple probe tip to the component under
test. Keep the probe touching the part until the reading
stabilizes (about 30 seconds).
Read the temperature on the display (F).
NOTICE
For longer period of measurements, use a thermal
tape to attach the thermocouple probe to the surface being
measured.
WARNING
temperature probe before changing to another measurement
function.
Risk of electric shock! Never touch the
Risk of electric shock! Remove the
EN
25