Télécharger Imprimer la page

AMAZON COMMERCIAL B083V7S41G Mode D'emploi page 20

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 34
EN
Press the PEAK/INRUSH button (L) for 2 seconds to return
to normal operation. The MAX/MIN indicator (8) no longer
appears.
VFD mode (Variable-Frequency Drive)
NOTICE
provide a more stable reading when testing variable frequency
motor drives.
NOTICE
In AC voltage measurement mode, press and hold the MODE/
VFD button (H) for 2 seconds. The VFD indicator (10) appears
on the display.
Connect in parallel the test probes (S) in parallel to the
device being measured.
Read the voltage on the display.
NOTICE
to return to normal operation. The VFD indicator (10) no longer
appears on the display (F).
DC voltage measurement
Insert the black connector lead (S) into the COM (black)
input terminal (I) and the red connector lead (S) into the V
(red) terminal (M).
Set the function switch (K) to the
20
The VFD mode filters high frequency noise to
The VFD mode range is 100V to 600V AC.
Press the MODE/VFD button (H) for 2 seconds
position.

Publicité

loading