4.

INSTALLATION: Triango 80 W

4.1

Workload data

Bending moment M
B
Vertical weight F
G

4.2. Installing the wall holder

Fastening hardware does not come with delivery.
DANGER
Assembly by qualified personnel
Assembly must be done by qualified personnel only.
Lack of appropriate knowledge may pose a death
hazard.
Two people are needed for assembly.
CAUTION
Determine the fasteners to use from the workload
data table
Before assembly, make sure the rod dimensions are
correct.
CAUTION
Make sure the wall holder is in the correct position.
The wall holder must be aligned according to the axis
in the image.
Misalignment will compromise mechanical safety.
We recommend using a counterplate on lightweight
construction walls (not included with delivery).
DANGER
Death hazard from electric shock.
All poles of the luminaire must be disconnected from
the power supply by an external lockable switch.
Draw 6 drill marks.
Observe the position of opening "A" for connecting
the current.
30
275 Nm
135 N
A
WARNING
Wear
safety
equipment
manufacturer's instructions.
Drill holes and blow out with a bellows.
Check the hole distances.
Position the wall bracket on the wall and hammer in the
fastening anchor.
Tighten the fastener according to manufacturer's
instructions.
DANGER
Death hazard from electric shock.
Do not turn on the power supply before the luminaire
head has been installed.
This device must only be connected to a power grid
with a protective ground conductor (PE) to prevent
risk of electric shock
Connect the power.