Instructions for Use
®
marCut
bipolar scissors
Cleaning: Mechanical
procedure
Disinfection
Drying
Maintenance, checks
and functional
inspection
Packaging
Sterilization
V. 4
Place instruments into a tray basket on the insert module or onto the inserts of
the MIS unit and start the cleaning process.
1 min pre-rinse with cold water
Emptying
3 min pre-rinse with cold water
Emptying
5 min washing at 55°C (131°F) with 0.5% alkaline
or at 45°C (113°F) with enzymatic cleaning agent
Emptying
3 min neutralization with warm mains water
(>40°C / >104°F) and neutralizer
Emptying
2 min intermediate rinsing with warm mains water (>40°C / >104°F)
Emptying
Perform thermal machine disinfection, observing all national requirements
regarding the A0 value (see ISO 15883).
Drying of the outside of the instruments by means of the drying cycle of the
cleaning/disinfection unit.
If necessary, the instruments can also be manually dried using a lint-free cloth.
Dry cavities of instruments using compressed air
Prior to each use, inspect the bipolar scissors for damage.
Withdraw damaged instruments from service immediately.
Use certified sterilization packing (e.g. according to EN 868, ISO 11607) or
sterilization containers.
Prior to sterilization, the instruments must be cleaned, disinfected, rinsed with
distilled water and finally dried carefully.
Method
Number of
pre-vacuum
phases
Fractionated
3
pre-vacuum
at a minimum
process (acc. to
of 60 mbar
ISO 13060/
ISO 17665)
As regards sterilizer handling, loading and parameter setting, please follow the
recommendations of the manufacturer of your machine.
As of 12/2014
.
Exposure
Exposure
temperature
time
min. 132°C
min. 3 min
(270°F)
max. 137°C
(278°F)
Drying
time
min.
10 min