The service must only be signed off in the maintenance plan if a minimum of all the above-
mentioned aspects have been checked.
Requirement:
1.) Check:
1.1) Power lead
EPR cable or comparable quality
1.2) Condition of the power lead
In particular look for:
crushing, abrasion, cuts, visible insulation
of the inner conductor, visible metallic veins, kinks,
bulges, color changes to outer cover,
brittle spots
1.3) Condition of internal cables
In particular look for:
crushing, abrasion, cuts, visible insulation of the inner
conductor, visible metallic veins, kinks, lumpiness, color
changes of the outer sleeve, brittle spots)
1.4) Sufficient ease of pulling and anti-kink of the
power cable
1.5) Sufficient ease of pulling and anti-kink of the
internal cables
1.6) The power cable and other electric leads must
be laid down securely in such a way that
shearing, squashing, and other mechanical
stresses are improbable.
1.7) Humidity protection IP X4
1.8) Device for suspending the power cable during
transportation
1.9) Check the frame parts for plastic deformation
and/or wear and tear.
- bottom of bed
- bed lift
- backrest, seat area, feet area
- boom
- boom socket
- rollers
1.10) Housing
Visually check all housings for damage, screws must
be securely fixed, seals and gaskets should not exhibit
visible damage.
Interval, Interval 120cm, Interval 140cm
Remark:
NA
Yes
NA
Yes
NA
Yes
NA
Yes
NA
Yes
NA
Yes
NA
Yes
NA
Yes
NA
Yes
NA
Yes
NA
Yes
!"#$%&.%
2011-01
No
Remark
No
Remark
No
Remark
No
Remark
No
Remark
No
Remark
No
Remark
No
Remark
No
Remark
No
Remark
No
Remark
EN

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Interval 120cmInterval140cm

Table des Matières