Haag-Streit BX 900 Mode D'emploi page 9

Lampe à fente photo
Masquer les pouces Voir aussi pour BX 900:
Table des Matières

Publicité

• The retinal dose for potential photo chemical damage de-
pends on ray density and examination time. When the ray
density is cut in half, the examination time doubles until the
border value is reached.
• Until today, there is no proof established that slit lamp
examination can endanger the eye. Nevertheless, we recom-
mend to use the minimum amount of light possible for retinal
examination.
• Children, Aphakics and people with eye problems are espe-
cially endangered. If the retina is exposed to the same or
another light source in 24 hours, there is also a higher risk.
This is especially true when a retinal picture is taken with flash.
• Only mirrors with LOT number may be used.
Warranty / product liability
• The instrument should be operated in accordance with the
chapter 'Safety'. Incorrect operation can damage the instru-
ment. Thus no warranty claims can be accepted.
• Continued use of an instrument that has been damaged by
incorrect operation can lead to personal injury. The manufac-
turer cannot accept liability in this case.
• Repairs and alterations on this equipment should only be
carried out by Haag-Streit service technicians or by authori-
zed persons.
Statutory requirements
• The slit lamp BX 900
®
has been designed to conform with the
IEC / EN 60 601-1, EN ISO 10939 and EN ISO 15004 stan-
dards. Manufacturing procedures, testing, commissioning,
maintenance and repair are conducted under the observance
of Swiss and international regulations.
• When combining different medical and/or nonmedical elec-
trical equipment standard IEC / EN 60 601-1 applies.
• The 'CE' marking confirms compliance of the slit lamp BX
900
®
with the directives 93/42/EEC and 89/336/EEC and the
conformity module A.
• A copy of the declaration of conformity of the present instru-
ment can be requested at any time at Haag-Streit company.
• All statutory accident prevention regulations are to be observed.
• Classification
IEC / EN Standard 60 601-1 Slit Lamp BX 900
safety class I.
Application part Type B.
Operation mode: continuous
operation
CE-Regulation 93/42 EEC
Class I
FDA
Class II
Pictograms
• Read very carefully (in the instruction manual)!
• Warning for hot surfaces!
• Disconnect the electrical supply plug!
Flash Unit FU 01
Lampengehäuse / Boîtier de la lampe / Lamp cover
Typenschild / Plaque signalétique / Legend plate
1)
Hersteller / Fabricant / Manufacturer
2)
HS-Part Number
3)
Gerätenummer / Numéro d'appareil / Equipment number
®
equipment
4)
Produkteklassifikation Typ B / Classification de produit
type B / Type B equipment
5)
Herstellungsdatum / Date de fabrication / Date of manufacture
• Protected earth!
• High voltage!
1500.7200646.04010
• Anschluss Auslösetaste
• Raccordement déclencheur électrique
• Connection shutter release bar
• Nach dem Ausschalten der Flash Unit FU 01
mindestens 15 Sekunden warten, bis das
Blitzkabel ausgezogen werden darf!
• Attendre au moins 15 secondes après le
déclenchement de la Flash Unit FU 01 avant
retirer le câble flash!
• After shut down of the Flash Unit FU 01 you
must wait at least 15 seconds until cable may
be removed!
• Anschluss Blitzkabel
• Raccordement câble flash
• Connection flash cable
• Nach dem Ausschalten der Flash Unit FU 01
mindestens 15 Sekunden warten, bis der
Schieber geöffnet werden darf.
• Attendre au moins 15 secondes après le
déclenchement de la Flash Unit FU 01 avant
de bouger le coulisseau!
• After shut down of the Flash Unit FU 01 you
must wait at least 15 seconds until the blo-
cking slide may be removed!