The filter is produced with (PTFE) hydrophobic material to prevent fluids entering the pneumatic circuit.
It should be changed immediately if it becomes wet or if there is any sign of contamination or discolouration.
If should also be changed if the unit is used with a patient whose risk of contamination is unknown.
Don't use the suction unit without the protection filter. If the suction unit is used in an emergency or in a patient where the risk of
contamination is not know the filter must be changed after each use. Available under request with different versions with complete jar 2000cc.
Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use.
Reuse may cause cross-infections. Don't use after lapse of the sell-by date
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1
standards on material biocompatibility.
The NEW ASKIR 30 suction equipment does not need maintenance or lubrication.
It is, however, necessary to inspect the unit before each use. Unpack the instrument and always check integrity of plastic parts and feeding
cable, they might have been damaged during previous use. Connect the cable to electrical network and turn the switch on.
Close the aspirator outlet with your finger and with suction regulator at maximum check that the vacuum indicators reaches at least -80kPa
(-0.80 bar). Rotate the knob from right to leff. The vacuum indicator should go down -40kPa (-0.40 bar).
Check that no loud noises are present. A protection fuses (F 1 x 1.6 A 250V) reachable from exterior and situated in the plug protects the
instrument. For fuses replacing, always the type and the range.
Type of fault
1. The suction unit doesn't work
2. No aspiration
3. No aspiration
4. The Vacuum power on the patient
side is either very low or absent
5. The float doesn't close
6. The float doesn't close
7. Low suction
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7
If the overfill security system fails, continued aspiration will be stopped by the antibacterial filter.
Should this back-up security system also fail, there is a risk of penetration of liquid into the device. Don'0t attempt to proceed with aspiration under these
circumstances.
Return the device to CA-MI for service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT CA-MI
TECHNICAL SERVICE.
CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE
TAMPERED.
ACCESSORIES SUPPLIES
DESCRIPTION
COMPLETE ASPIRATION JAR 1000cc
CONICAL FITTING
TUBES SET 6mm x 10mm (TRASPARENT SILICON)
ASPIRATION PROBE CH 20
ANTIBACTERIAL FILTER

PERIODICAL MAINTENANCE CHECKS

Cause
Cable is damaged
External power source failure
Jar Cap not screwed on correctly
Lid seal not in its seat
• Vacuum regulator set to minimum
• Protection filter blocked or damaged
• Connection tubes blocked, kinked or
disconnected
• Shut-off valve blocked or damaged
• Pump motor damaged
If the cap has been washed, ensure that
the float is not partially detached
The float it's covered by dirty material
Foam inside the jar
None of the procedures have achieved
the desired results
Remedy
Replace the cable
Check the external power source
Unscrew the cap, and re-screw it
Unscrew the cap and insert the seal properly in its seat
• Turn the vacuum regulator clockwise and check the
value of the vacuum on the gauge
• Replace the filter
• Replace or reconnect the tubes, check the jar
connections
• Empty the jar, or disconnect the tube from the jar and
unblock the shut-off valve. The unit twill only work in
the upright position
• Refer to authorised service personnel
Fit the float into it's place
Unscrewed the cap, leave the and put in on autoclave
Fill the jar to 1/3 full of ordinary water
Contact CA-MI customer service