Chapter 4 - Technical Specification

4.1.6 Table requirements

Do 
NOT
plugged directly into a wall socket.
The scan head should be positioned in the middle of the table width. The front of the scan
WARNING
head should be inline with the front of the table to allow the patient to get into the correct
alignment position.
The scan head and touch screen must be positioned on a table that meets the necessary safety standards. The
table
support the weight of the scan head.
MUST

TABLE 8: Table size requirements

Dimension
Details
Width
700 mm (28 inches)
Depth
550 mm (22 inches)
Height
645 mm (25.4 inches) to 945 mm (37.2 inches)
Where wheelchair accessibility is required the table must meet wheelchair accessibility standards.

4.1.7 Battery

Do
NOT
permitted to service the device.
CAUTION
The device contains a 3V Lithium coin battery.

4.1.8 Environmental considerations (ambient)

operate the device in an environment that exceeds the environmental limits, see Warnings and
Do
NOT
cautions on page 11 .

TABLE 9: Environmental ranges

Status
Normal use
Storage
Transport
Note
Device is DEHP free.
Page 54 of 65
English
plug the scan head into a power block or powered table. The scan head must be
remove the scan head covers. Only Optos trained technical personnel are
Temperature
+10°C to +35°C
-10°C to +60°C
-10°C to +60°C
Relative Humidity
(Non-condensing)
30% to 75%
10% to 95%
10% to 95%
Copyright 2017, Optos plc. All rights reserved.
Atmospheric Pressure
700hPa to 1060hPa
700hPa to 1060hPa
500hPa to 1060hPa
Part Number: G102748/9GWE