Introduction; Device Description; Intended Use - Nikon OPTOS Daytona P200T Manuel D'utilisation

Ophtalmoscope panoramique
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 202
2

Introduction

Welcome to your User Guide. This guide covers the following products:
P200T
l
P200T
l
Note
In this guide the term P200T is used when referring to all product variants, unless otherwise stated.

2.1 Device description

This confocal scanning laser ophthalmoscope is a widefield digital imaging device capable of capturing the area
from the central pole to the far periphery of the retina. The retinal images are captured in an automated, patient
friendly way without scleral depression or contact with the cornea.
The P200T uses red and green lasers to produce a digital, high-resolution image, which is displayed on a PC
monitor screen.
plus
The P200T
and superior areas of the opto map image.
Widefield Confocal Scanning Laser Ophthalmoscope (SLO) fundus imaging is achieved using red and green
lasers. These laser wavelengths penetrate the retinal structures to different depths. Each wavelength provides
information for interpretation and diagnosis.
In autofluorescence mode, the device captures opto map
This allows an image of the natural fluorescence of the eye to be captured. Fluorescent dye is not introduced
into the patient when performing autofluorescence imaging.
Optos' technology is designed to operate through a minimum pupil diameter of 2 mm. Although pupil dilation is
required, the decision to dilate is a medical decision to be made by the eyecare professional.
NOT
Images are captured on the scan head and then viewed, magnified, annotated and separated into their
components in the review software.
Note
Depending on the features available on your device, you may have a subset of the features described in this
user guide.

2.2 Intended use

The P200T scanning laser ophthalmoscope is intended to be used as a widefield and retinal autofluorescence
imaging ophthalmoscope to aid in the diagnosis and monitoring of diseases or disorders that manifest in the
retina.
Part Number: G102748/9GWE
Copyright 2017, Optos plc. All rights reserved.
plus
differs from the standard P200T in that it introduces enhanced image quality in both the inferior
af
images using the green laser to illuminate the eye.
Chapter 2 - Introduction
Page 21 of 65
English

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Optos daytona p200t plus

Table des Matières