Chapter 3 - How to...
Cleaning before each patient
The patient should not touch the touch screen. The patient should only be in contact with
the face pad, eye piece and hand grips.
Always follow the cleaning and disinfecting agents' manufacturer's instructions and
precautions; for storage, method of application, protective gloves, shelf life and disposal
to help assure effective and safe use of the product.
Do
NOT
CAUTION
piece and hand grips) in water or cleaning fluid.
Do
NOT
scan head. Always allow cleaned areas to air-dry.
Do
NOT
The following cleaning procedures should be carried out before each session.
1. Clean the eye piece and dust cap with an individually sealed, 70% IPA (isopropyl alcohol) wipe. The
eye piece and dust cap can be removed for cleaning on both sides.
2. Clean the hand grips with an individually sealed, 70% IPA (isopropyl alcohol) wipe.
3. Allow the cleaned areas to air-dry before imaging the next patient.
Note
The wipes should be checked periodically to ensure that they are within their marked expiration date.
Cleaning the main mirror
The mirror can be damaged during cleaning. Preventing mirror contamination reduces the
risk of damaging the mirror. It is
Only clean the mirror if dust particles, smudges or spots have accumulated on its surface.
When cleaning the main mirror, do
Do
NOT
Only clean visible areas of the main mirror. Do
into parts of the device you cannot see.
CAUTION
Do
NOT
Do
NOT
surface as this may damage the mirror.
Dispose of the dirty cloth after each wipe to avoid transferring the dust back on to the
mirror. You will have to use several wipes to clean the mirror.
Depending on the operating environment dust may accumulate on the device that may affect image quality. To
clean the main mirror follow the procedure below:
Note
Points to remember when cleaning the main mirror:
To prevent contamination, take care
l
the mirror.
Only hold the cloth by the corners.
l
Page 44 of 65
English
attempt to wash or immerse the patient contact parts (for example, face pad, eye
use lint cloths, tissues, or other materials that may create dust, on or near the
let the cleaning wipes come into contact with the inside of the instrument.
NOT
touch the edges of the main mirror. The edges of the main mirror are sharp.
polish the mirror surfaces; clean only dust particles, smudges or spots.
wear jewelry (for example rings, watches or bracelets) while cleaning the mirror
NOT
to touch the surface of the cloth that will be used to clean
recommended to routinely clean the main mirror.
rub the spots too hard, as this may cause damage.
NOT
put your hand or any cleaning material
NOT