Chapter 4 - Technical Specification
Feature description
Power Consumption
The touch screen
be powered through the
MUST
scan head and is included in these figures.
Touch screen voltage

4.1.2 Connections

These connections are
NOT
FIGURE 5: Sample of the connection label
No. Description
Mains power cable and switch
1
Protective earth must be fitted.
l
U.K. Power Cable Fuse: 13A
l
2
Network cable
3
Touch Screen
4
USB. For Optos Service use only.
5
Audio 3.5mm jack - unused.

4.1.3 Scan head dimensions

The scan head and touch screen should be placed on a table which meets the table specification, see Table
requirements on page 54 . There should be adequate clearance around the table to operate the device.
The image server, switch, browser-based review clients and any Viewing PCs or monitors must be positioned
outside the patient environment; more than 1.5 meters (5 ft) from the scan head.
Page 52 of 65
English
Compliance
300VA
3A at 100-120Vac
1.5A at 200-240Vac
SMPS continuous
+12V
intended to be detached by the user.
Part Number: G102748/9GWE
Copyright 2017, Optos plc. All rights reserved.