Error
Cause
The red LED of the
The battery is defective.
charger flashes.
The green LED of
Power outlet carries
the charger fails to
no voltage.
light up, even though
the power cable is
connected to a power
outlet.
Charger is defective.
Maintenance and Care
Replacement of the Battery
Never place the handpiece in the charger without the
battery inserted in the handpiece!
Use 3M ESPE batteries only! The use of other manu-
facturer batteries or non-rechargeable batteries/
primary batteries is a potential hazard and may
damage the device.
Rotate the lid at the lower end of the handpiece
counterclockwise until it hits the stop, and then
remove the lid.
Push a small spatula or similar aid between the
cylinders of the battery until the snap-lock releases
so that the battery protrudes slightly from the hand-
piece body.
Remove the battery from the handpiece.
Solution
Replace the battery.
Use a different power
outlet.
Have the charger repaired.
Push the (replacement) battery into the handpiece
proceeding along the direction shown by the arrow
until the battery locks into place.
Replace the lid and close by rotating clockwise.
Insert the handpiece into the unit to completely
charge the new battery for its first use.
- The red LED of the charger illuminates after
approx. 2 sec; please see also "LED Display of
the Charger". The battery charging process is
complete once the red LED is turned off.
- Note: the new battery attains full capacity only
after several charging/discharging cycles.
Therefore, the battery charge may initially be
sufficient for a smaller number of exposures only.
Handpiece/Battery Care
The charging contacts of the charger and handpiece
must not be contacted by metallic or greasy parts.
Do not use any other chargers as this may damage
the battery!
The battery must not be immersed in water or
thrown into open fires!
Cleaning the Light Guide
The light guide can be autoclaved.
Do not sterilize by chemical means or hot-air!
The light guide should be regularly wiped clean
with a soft cloth. Especially before and after steam
sterilization the spots of dried liquid should be
wiped off the ends of the device.
Adhering polymerized composite should be removed
with alcohol, a plastic spatula may help in removing
the material.
- Do not use any sharp or pointed tools to protect
the surface of the device from scratching.
Cleaning the Charger and Handpiece
For disinfection of all components of the device,
spray the disinfection agent onto a piece of cloth
and proceed to disinfect the device.
- Disinfection agents must not enter the device!
- Pursept-A (from Merz) and FD 322 (from Dürr) are
recommended disinfection agents.
Clean the charger and the handpiece with a soft
cloth and a mild cleaning agent, if required.
- It is not permissible to use solvents or abrasive
cleaning agents as these may damage plastic
parts of the device!
- Cleaning agents must not enter the devices!