3M ESPE Elipar FreeLight Mode D'emploi page 7

Polymérisateur sans fil
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

If the light intensity is below 80% - fewer than
4 LEDs are ON - check the light guide for contami-
nation or defects.
Proceed as follows:
(1.) Clean the light guide, if contaminated; please
refer to the section on "Care" or
(2.) replace the light guide, if defective, or
(3.) if (1.) and (2.) fail to improve the situation, please
contact the 3M ESPE Customer Service or your
retailer.
Low Battery Charge Display
After frequent use the battery charge of the device may
drop to below 10%, so that only a few more exposures
can be performed without re-charging. Low battery
charge is shown by several means:
- A short audible alarm is emitted 5x at the com-
pletion of an exposure, and repeated thereafter
whenever any button is touched.
- The 10-sec-LED of the handpiece begins to flash.
The handpiece should be placed in the charger as
soon as possible to re-charge the battery.
Power-Down Mode
Once the handpiece is placed in the charger, all internal
functions and LEDs are automatically turned off as the
handpiece switches to power-down mode. This reduces
the power consumption of the rechargeable battery to
a minimal level. Outside the charger, the handpiece is
also switched to power-down mode if left unused for
approx.10 min.
To terminate the power-down mode press one of
the two buttons.
- The power-down termination signal (two short
audible signals) is emitted indicating that the
6
100%
80%
60%
40%
blue LEDs
20%
handpiece is ready for operation: the handpiece
displays show the latest selected exposure mode
and time settings.
Acoustical Signals
An acoustical signal is emitted
• every time a button is pressed,
• every time the light is turned ON or OFF,
• after 10 sec of exposure time have elapsed (1 alarm
signal), 2x after 20 sec, 3x after 30 sec.
Two acoustical signals are emitted
• every time the power-down mode is terminated by
pressing any of the buttons.
A 2 sec-error signal is emitted, if
• the handpiece over-heats,
• the rechargeable battery is discharged to an extent
that reliable polymerization is no longer ensured.
Operating Errors
Error
The 10 sec-LED
within the dark blue
frame flashes and
a short signal is
emitted 5x when-
ever a button is
touched or the light
is turned off.
The ongoing exposure The battery lacks sufficient
is interrupted (light
off signal is emitted)
followed by a 2 sec-
error signal; the device
switches to power-
down mode and resists
further activation.
A 2 sec-error signal
is emitted upon
pressing of the
Start button.
Cause
Solution
The residual battery charge
has dropped below 10%.
Place the handpiece in
the charger and re-charge
the battery.
charge.
Place the handpiece in
the charger and re-charge
the battery.
The handpiece has become
overheated in the course
of use.
The handpiece can be used
again once it has cooled
down.
Allow the handpiece to
cool down.
- The handpiece can be
used again, once pressing
the Start button success-
fully activates the light.