Error codes

Basic error codes:
ATTENTION! If an error code is encountered, an audio warning
signal is generated and the red Off signal light goes on in the dis-
play field. First, please try to continue operation by turning the
device off and back on again.If it is still not possible to eliminate
an error after an extended wait, please contact our service
department. If you do so, you should always tell us what error
code has been encountered. This simplifies the process of
tracking down the error and makes it possible to form a prelimi-
nary opinion.
032007
160
240
320
400
480
560
640
160
240
320
400
480
560
640
Error code
Off - LED (red)
1x flash /
2 sec pause
Off - LED (red)
2x flash /
2 sec pause
1/min
720
800
Off - LED (red)
3x flash /
2 sec pause
Off - LED (red)
4x flash /
2 sec pause
1/min
Off - LED (red)
720
800
5x flash /
2 sec pause
Off - LED (red)
6x flash /
2 sec pause
Off - LED (red)
7x flash /
2 sec pause
Off - LED (red)
8x flash /
2 sec pause
Error
Cause of the error
The potentiometer for
-Too rapid a change in the
the speed is not assig-
speed target value
ning any target value
-Internal device error
Permissible motor tem- -Permissible supported
peratur exceeded
weight exceeded
-Permissible ambient
temperature exceeded
Motor locked
-Agitation table is being
hindered in this stroke mo-
tion by an outside effect
Read fork signal not
-Internal device error
available
Maximum adjustable
-Resonance behavior of
speed of the device
the experiment setup
exceeded
Motor standstill for de- -When turning on the
vice switch-on proce-
device, you must ensure
(safety confirma-
that the motor is not
dure
tion not ensured
performing any rotational
motion
Safety relay cannot
-Internal device error
be activated
The switch for the TIME -Internal device error
function is not assigning
any target value
Problem with the power -The device is beeing
supply
operating with under vol-
tage or overvoltage (for
permissible voltage range,
see technical data)
Triac cannot be activated -Internal device error