Intended Use; Room Climate Classes - Vestfrost ULTF 40 VTS098 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN

Intended use

Vestfrost biomedical freezers are designed to provide reliable temperature stability and precision
and ensure an ideal cold temperature storage for sensitive items is achieved� The products are
designed for the following operating ranges:
Models
Temperature range
Fabric Set Point
Ambient Temperature
Relative humidity
Number of probes
WARNING: This unit is not a "rapid-freeze" device� Freezing large quantities of liquid,
or high-water content items, will temporarily increase the chamber temperature and will cause
the compressors to operate for a prolonged time period� Avoid opening the door for extended
time periods since chamber temperature air will escape rapidly� Also, keep the inner doors closed
as much as possible� When room air, which is higher in humidity, replaces chamber air, frost may
develop in the chamber more rapidly�

Room climate classes

The climate class is stated on the nameplate� This specifies the optimum room temperature�
ULTF40/ VTS098
VTS252
-60°C to -86°C
-15°C to -30°C
-82°C
+16ºC to +32ºC
+16ºC to +32ºC
MAX 70 %
MAX 70 %
1
Room climate
class
3
4
5
7
VTS254
-25°C to -45°C
-25°C
-45°C
+16ºC to+ 32ºC
MAX 70 %
1
1
Max room
Relative
temperature
humidity
°C
25
30
40
35
Vestfrost Solutions
VTS256
-40°C to -60°C
-61°C
+16ºC to +32ºC
+16ºC to +32ºC
MAX 70 %
1
%
60
55
40
75
9
VTS258
-60°C to -86°C
-82°C
MAX 70 %
1

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ultf 40 vts252Ultf 40 vts254Ultf 40 vts256Ultf 40 vts258

Table des Matières