Electrical connection

Wiring and connections in power supply systems must been all applicable (local and national)
electrical codes. Consult these codes lengths and sizes prior to cabinet installation.
This device complies with relevant EU directives including Low Voltage Directive 2014/35/EU and
Electromagnetic Compatibility Directive 2014/30/EU
The socket should be freely accessible.
Connect the appliance only to 220-240V / 50Hz mains supply via a correctly installed earthed
socket.
The socket must be fused with a 10A or 13A fuse.
If the appliance is to be operated in a non-European country, check on the rating plate whether
the indicated voltage and current correspond to the values of your mains supply.
Data regarding voltage and obsorbed power / current are given on the rating plate.
The power cord may be replaced by a technician only.
The rating plate provides various technical information as well as type and serial number.
EN
1-NXX866400710590
Product number
EN

Intended use

Vestfrost biomedical freezers are designed to provide reliable temperature stability and precision
and ensure an ideal cold temperature storage for sensitive items is achieved. The products are
designed for the following operating ranges:
Models
ULTF-37i
Temperature range
-60°C to -86°C
Fabric Set Point
-82°C
Ambient Temperature
+16ºC to +32ºC
Relative humidity
Max 70 %
Number of probes
1
WARNING: This unit is not a "rapid-freeze" device. Freezing large quantities of liquid,
or high-water content items, will temporarily increase the chamber temperature and will cause
the compressors to operate for a prolonged time period. Avoid opening the door for extended
time periods since chamber temperature air will escape rapidly. Also, keep the inner doors closed
as much as possible. When room air, which is higher in humidity, replaces chamber air, frost may
develop in the chamber more rapidly.

Room climate classes

The climate class is stated on the nameplate. This specifies the optimum room temperature.
Room climate class
Max room temperature
3
4
5
40
7
35
Vestfrost Solutions
Vestfrost Solutions
Relative humidity
°C
%
25
60
30
55
40
75
9
9