Voltage stabilizers

It is highly recommended that all cold chain
equipment running on AC current is installed
with a voltage stabilizer, in order to safeguard
the equipment against voltage fluctuations.
Stabilizers should be selected & installed as
per the input voltage available.
Instructions to User:
● Every refrigeration unit must be connected
to an individual stabilizer.
● Bypassing of Stabilizer is not recommend-
ed, as such practice may lead to damage
of the CCE & in turn safety of vaccines &
hence must be avoided.
● Proper earthing should be available and
connected.
● Emphasize on repairing stabilizers immedi-
ately. Identify authorized Vestfrost service
provider in case of technical assistance is
required.
The voltage stabilizer is also provided with ar-
rangement to cut off its output voltage to the
Vaccine/waterpacks freezer in case the mains
voltage goes below or above as per designed
input voltage range. The output is restored
automatically after the factory set delay 3 to 6
minutes (WHO/PQS E007/VS01.4 specifica-
tions) when the mains voltage is within the
recommended range and remains within.

Electrical connection

Wiring and connections in power supply
systems must been all applicable (local and
national) electrical codes. Consult these
codes lengths and sizes prior to cabinet
installation.
This device complies with relevant EU
directives including Low Voltage Directive
2014/35/EU and Electromagnetic Compat-
ibility Directive 2014/30/EU
The socket should be freely accessible.
Connect the appliance only to 220/240 V
50Hz alternating current via a correctly
installed earthed socket.
The socket must be fused with a 10-13 A
fuse.
If the appliance is to be operated in a non-
European country, check on the rating plate
whether the indicated voltage and current
type correspond to the values of your mains
supply.
Information regarding voltage, current or
power are given on the rating plate
The power cord may be replaced by a
technician only.
Check that the plug fits your type of socket.
If it does not fit, have a qualified electri-
cian to fit the plug. This appliance must be
earthed.
Connect the appliance to voltage stabi-
lizer and connect voltage stabilizer to wall
socket. (Fig 4 + 5)