USER CARE AND MAINTENANCE
The unit should be kept clean and free from moisture and dust. Clean the unit at least weekly by wiping with a
dry, lint-free cloth. Avoid getting fluids or debris such as sand or dirt inside the oxygen connections. Do not
immerse in water. Do not clean with a solvent based cleaning solution. Avoid dropping the unit or placing it in a
position where it could topple or fall since this can damage the device. Whenever possible, use a padded
carrying bag such as those listed in the Accessories section to carry the unit. This will help to protect it in case
of a fall. The unit should be protected from extreme temperatures. Do not attempt any other maintenance.
STORAGE & hANDLING WARNINGS
When attached to a cylinder, do not place oxygen cylinders in unventilated spaces such as car trunks.
Excessive heat can make the relief valve suddenly and quickly discharge the cylinder contents,
possibly making it a projectile and greatly increasing the oxygen level in unventilated spaces.
Do not leave oxygen cylinders in the cabs of vehicles without ventilation. If a cylinder leaks, a spark
could start a fire causing serious injury or death.
Remove cylinders from the vehicle when the destination is reached. Be sure to secure cylinders from
movement during transport.
Heat, Humidity, Sun and Artificial Light have no effect on operation as long as the device is used
within the product specifications.
This device contains electrical and/or electronic equipment. Follow local governing ordinances and
recycling plans regarding disposal of device components.
TROUbLEShOOTING
WARNING
Do not attempt to open the device for maintenance or repair. The device contains no user-serviceable
parts. Do not attempt any other maintenance. Contact your oxygen provider if service is required. If
you do not have a healthcare provider, refer to the DeVilbiss contact information on the back of this
guide.
SYMPTOMS
Oxygen is not being delivered
even though the PulseDose
indicator is flashing every
time I breathe.
Use times are different from
those stated in the literature.
A-1000
POSSIBLE CAUSES
1. Oxygen supply is empty.
2. Oxygen supply is not turned
on.
1. PulseDose responds to your
breath rate. Your breath rate
may vary, which causes the
operation time to vary.
2. Leak in system
REMEDIES
1. Check contents indicator on the device.
If empty, switch cylinders.
2. Open the compressed oxygen cylinder
valve by following the directions given by
your service representative.
1. PulseDose is operating correctly.
2. Check connection to cylinder. It may
need a new regulator seal.
EN - 9