Cleaning, Disinfection, Sterilization - Dentsply Maillefer wave-one Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

6.9.

Cleaning, Disinfection, Sterilization

WARNINGS
 The components need to be disinfected and/or sterilized (only for contra-angle) prior to first
use and in between each case!
 Do not place the micromotor or any other accessories in an autoclave unit or ultrasonic
tank. Only the contra-angle may be sterilized (see separate contra-angle operating
instructions).
 Do not immerse the device in ultrasonic cleaners.
 The plastic enclosure is not sealed, do not use any liquid or spray directly on the console,
especially on the monitor or near the electrical sockets.
Exterior Surfaces
Wipe the surface of the device (panel interface and the housing), the micromotor and its cable
with a clean cloth moistened lightly with a non-aggressive disinfectant.
Use only disinfectants with antibacterial, fungicide and viricide power and which comply with
the valid requirements of the national regulatory body.
It is recommended to use aldehyde-free disinfectants.
Contra-angle
To sterilize the WaveOne
operating instructions.
Endodontic Files
To sterilize the endodontic files, refer to the manufacturer's instructions for use.
(MAILLEFER 6:1) contra-angle, refer to the separate contra-angle
www.enversaltan.com
www.enversaltan.com
- 24 /141 -

Publicité

Table des Matières
loading

Table des Matières