D. SAMPLE WEIGHT IN WATER
Then the display shows "0.00g in-L" .Hang the sample on the hook (If a con-
tainer is used, hang the container on the hook and put it into the water, then
press TARE key to subtract the container's weight first), and then put the sam-
ple in the water. The display will show a weight value, which indicates the
sample's weight in water.
in-L: Indicate weight in water
The display shows as below
Press
E. Then the display will show a value which indicates the density of the sam-
ple. The density calculation is finished. (ρ-O: density of the sample)
For example: "33.6841r-O"
Press
Press
weighing mode.
REMARKS
1. The density unit is g/cm3, so if the weighing unit is not "g" when entering into
this mode, it will automatically be changed to "g" . And when exiting from this
mode, the weighing unit will be back to the original one.
2. Press
3. Zeroing, Tarring and cancelling tare functions are available in this mode.
4. Density calculation formula:
5. Press
water temperature will be the same as the last one. But if exiting from this
mode, the water temperature will be the default one when entering into the
mode for the next time
12
key to confirm
key to confirm and do the density calculation of next object.
key to exit from the density calculation mode and return to
key to exit from this mode at any time.
key in step "e" to do the density calculation for next object, the
USER MANUAL BAP
r-O
Density of the sample
r-L
Density of the water
Sample weight in air
W
A
Sample weight in water
W
L