WALL MOUNTING
Always cut off the power supply at the installation site prior to installation
- After having fixed as reference point the chair centre (see fig. A-B), carry out on the wall three
d.14 bores at the bores of the wall mounting (3); pay attention the perpendicularity of the bore.
- Insert the expanders (12) in the d.14 bores previously carried out and block with the special
hex spanner (installation accessory) the 2 screws; do not choke the cable between the application
(3) and the wall.
- Apply three plugs (1) on the wall mounting (3) bores.
- Unscrew the screw (8) Remove the cover (7), insert the lamp arm inside the application and
grease the pin. Connect the lamp cables to the terminal board (5) (see the electrical diagram
below), including the earthing cable. Connect the wires which come out of the wall to the
terminal board, if previously walled in. In the absence of such arrangement,
the connection as to be carried out with a movable cable, to be passed in the cable gland (10).
- Fit back the cover (7) by means of screws (8).
WIRING DIAGRAM
Wall mounting
Light cables
fig.A
fig.A
Terminal board (5)
fig.B
fig.B
Mains cable earthing connection
Power supply
12-24 Vac
17-33 Vdc
53 53