Power supply
4 x 1,5 V
Battery life
For approx. 220 measurements, de-
pending on the blood pressure level
and/or pump pressure
Accessories
Instruction for use, 4 x 1.5 V AAA bat-
teries, storage pouch
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
Technical information is subject to change without notifica-
tion to allow for updates.
• This unit is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to
electromagnetic compatibility (EMC). Please note that
portable and mobile HF communication systems may in-
terfere with this unit. More details can be requested from
the stated Customer Service address or found at the end
of the instructions for use.
• This device is in line with the EU Medical Devices Direc-
tive 93/42/EC, the „Medizinproduktegesetz" (German
Medical Devices Act) and the standards EN 1060-1
(non-invasive sphygmomanometers, Part 1: General
requirements), EN 1060-3 (non-invasive sphygmoma-
All manuals and user guides at all-guides.com
AAA batteries
27
nometers, Part 3: Supplementary requirements for
electro-mechanical blood pressure measuring systems)
and IEC 80601-2-30 (Medical electrical equipment –
Part 2 – 30: Particular requirements for the safety and
essential performance of automated non-invasive blood
pressure monitors).
• The accuracy of this blood pressure monitor has been
carefully checked and developed with regard to a long
useful life. If using the device for commercial medical pur-
poses, it must be regularly tested for accuracy by appro-
priate means. Precise instructions for checking accuracy
may be requested from the service address.