<When sterilizing components except for a Tip>
Type
Minimum Temperature
Full Cycle Time
Minimum Dry Time
Do not exceed 138˚C during sterilization.
CAUTION
4) Keep the components in the autoclave pouch until ready to use.
5) Store in a dry and dust-free location up to the period specified by the sterilization pouch manufacturer. If sterility cannot be
confirmed, sterilize again prior to use.
• Do not sterilize by ultraviolet ray. The handpiece could discolor.
• If autoclaved with other instruments stained with chemical solution, it could strip the plating
CAUTION
and make the surface black.
• Do not autoclave any parts (the Control Unit, AC Adaptor, Foot Control, Handpiece Cord,
Tip Card, Tip Power Guide, O-Ring). Other than those that can be subjected to autoclave
sterilization. Perform alcohol disinfection to the Control Unit, AC Adaptor, Foot Control,
Handpiece Cord, Tip Card, Tip-Power Guide including after every patient.
• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
• Diamond coated Scaler Tip is for single use only. Do not sterilize and use again.
• Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and
sunlight. The air should be free from dust, salt and sulphur.
• Do not touch the product immediately after autoclaving may cause injury.
Steam Sterilization Cycle
Gravity Displacement
121˚C
132˚C
20 minutes
15 minutes
10 minutes
10 minutes
Pre-vacuum
135˚C
135˚C
10 minutes
10 minutes
10 minutes
15 minutes