Mem Programs - I-Tech MIO-IONOTENS Notice D'utilisation

Masquer les pouces Voir aussi pour MIO-IONOTENS:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Position of electrodes: from photo 1 to photo 20 of the Positions manual.
Intensity: Apply with caution (at low intensity, enough to produce light muscle
contractions) in the first 2/3 weeks. Increase intensity progressively over the
next 3/4 weeks.

MEM programs

Medical
Prg
prg.
Yes/No
1-5
Yes
6-10
No
11-
No
12
13
No
M1-M5 • TENS Free memories (medical program)
Free memories for antalgic TENS treatment.
M6-M10 • NEMS Free memories (non-medical program)
Free memories for muscle recovery and training.
M11-M12 • NEMS Alternated free memories (non-medical program)
Free memories for muscle recovery and/or training with alternated impulses
on channel 1 and 2.
M13 • Battery test program (non-medical program)
Battery calibration program for the exclusive use of the manufacturer.
IACER Srl
Description
Free
memoriesTENS
Free memories
NEMS
Free memories
NEMS alternated
CH1/CH2
Battery test
77
PHASE 1
Total time 1-90 min
frequency 1-200 Hz
width impulse 20-250 µs
Total time 1-90 min
frequency 1-200 Hz
contraction time 1-10 sec
slope 0-5 sec
recovery time 0-30 sec
width impulse 50-450µs
Total time 1-90 min
frequency 1-200 Hz
contraction time 1-10 sec
slope 0-5 sec
recovery time 0-30 sec
width impulse 50-450µs
MNPG151-07