Short operating Instructions Dräger FG4500

General safety instructions

Any use of the Dräger FG4500 requires full understanding and compliance with
the operating instructions 5695100, the relevant standards, as well as the relevant
statutory regulations.
The device is intended only for the uses described in this manual. Any improper use of the device may lead
to electric shock or destruction of the measuring instrument!
Before using the measuring device for the first time, the integrity of all components must be checked, e.g.:
Device exhibits no visible damage
No condensed water in the gas preconditioner
The gas preconditioner filter is clean
Gas hoses without defects
Visual inspection of the probe
Always fully charge the Dräger FG4500 via the USB interface with a 5 V DC / 1,5 A USB
power supply adapter only.
Incomplete charging affects the charging capacity of the battery in the long term.
When charging, no measurements should be performed.
O perating Instructions and managing PC measuring data
The operating instructions and the measurement data management can be found on our website
www.draeger-msi.de/home/?L=1
necessary USB drivers are installed automatically by downloading the PC data management.
Since 2005, EU-wide regulations apply to the disposal of electrical and electronic equipment. In
essence, these regulations set out that collection and recycling options must be available for private
households. Because the Dräger FG4500 is not registered for use in private households, it must not
be disposed of through such channels. The devices can be returned to your national retailer or to your
national Dräger Safety Organization for disposal. Please contact Dräger MSI GmbH, if you have any
questions regarding disposal.
Maintenance
To ensure proper function and accuracy, calibration and adjustment should be performed once a year.
Device maintenance must only be carried out by trained service personal.
Switching on
Press the On-/Off Button slightly for 1 second, then press Next.
Switching off
Select and activate the OFF menu item in the main menu or press the the On-/Off Button for 5 seconds
whilst in any menu point.
Make sure that the gas outlet on the side is unobstructed and not closed or blocked!
Functional areas
The individual tests and measurements are accessed via the main menu.
Flue gas measurement
To perform a complete flue gas measurement, we recommend a measurement time of at least 2 minutes. In
addition to the flue gas measurement, it is also possible to calculate averages and measure the draft and
enter the combustion system date.
Switch on the device and wait for system check. Then connect the flue gas probe to the unit (see connection
diagram).