Commissioning
3.4 Transducer Installation
Figure 3-8

Final Setup

1. At flowmeter, scroll down to [Install Complete]. Press the <Right Arrow> key.
2. Press the <Down Arrow> key and select [Install].
3. Press <ENTER> key. Flowmeter will go through drives.
4. Observe the Measured Vs window and verify a correct sound velocity measurement (if
5. Press the <Down Arrow> to accept sound velocity value.
6. The fllowmeter is now ready to report flow.
26
Single Channel
Dual Channel
Dual Channel
Channel 2
Channel 1
To Downstream Transducer
Connecting Transducers to Flowmeter
known).
To Upstream Transducer
To Downstream Transducer
To Upstream Transducer
Single Channel
To Downstream Transducer
To Upstream Transducer
SITRANS FUS1020 IP65 NEMA 4 Quick Start
Operating Instructions (Compact), 4/2009, CQO:QSG006 Revision B