Select Liquid Class
1. Press <Left Arrow> key to select [Pipe Data] and then press <Down Arrow> key to select
2. Press <Right Arrow> key to select [Liquid Class].
3. Press <Right Arrow> key again to select [Select Liquid]. Press <Right Arrow> key again.
4. Press <Up/Down Arrow> keys to scroll to desired liquid.
5. Press <ENTER> key to save selection.
Colon appears.
Select new liquid from list.
Select Pipe Configuration
1. Press <Left Arrow> key and then <Down Arrow> key to select [Pipe Config].
2. Press <Right Arrow> key.
SITRANS FUS1020 IP65 NEMA 4 Quick Start
Operating Instructions (Compact), 4/2009, CQO:QSG006 Revision B
Note
Pre-programmed Pipe Size and relevant pipe parameters will appear in menu cells. To
enter dimensions manually if pre-programmed dimensions do not match application, refer
to flowmeter manual for details.
Colon must appear first before Pipe Class can be selected.
[Application Data].
Commissioning
3.3 Setting the Parameters
19