Description And Intended Use; Intended Use; Cautions And Warnings; Limited Warranty - Laerdal LCSU 4 Manuel D'utilisation

Masquer les pouces Voir aussi pour LCSU 4:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 17

Intended Use

The LCSU 4 is a portable, electrically powered, medical suction device intended for field and transport use. It is intended for
intermittent operation to remove secretions, blood or vomit from a patient's airway to allow ventilation. Higher vacuum levels are
generally selected for oropharyngeal suction, and lower vacuum levels are usually selected for tracheal suctioning and the suctioning
of children and infants.
Notes
Do not use this unit until you have carefully read and fully understood these User Guides. Contact Laerdal Medical or its
authorized distributor for additional information if required.
Federal law (USA) restricts the LCSU 4 to sale by or on the order of a physician or other licensed medical authority.
Use only Laerdal accessories supplied directly by Laerdal Medical or one of its authorized distributors to ensure that the
LCSU 4 operates satisfactorily.
When using the LCSU 4 at home, take care to keep out of reach of children and pets. Do not store the LCSU 4 next to a
heater.

Cautions and Warnings

Cautions
The LCSU 4 is not suitable for use in the presence of flammable liquids or gases; danger of explosion or fire.
Do not use the LCSU 4 under environmental conditions that are outside ranges specified. This can endanger safety and
adversely affect device operation.
Ingress of suctioned material into the pump can damage and/or disable the device. If suctioning of liquid from the Canister or
patient into the pump is suspected, do not use the LCSU 4. Contact Laerdal Medical or your authorized distributor for advice.
Changes or modifications not expressly approved by Laerdal Medical could void the user's authority to operate the equipment.
Warnings
The LCSU 4 should only be used by persons trained in the use of medical suction equipment, and according to local protocol.
Unauthorized service attempts, opening or tampering with the LCSU 4 or its electrical components can damage or disable the
device, and will void the Limited Warranty.
Cat. No. 880052/880062 are approved according to RTCA/DO-160G Section 21 Category M but limited to battery operation
use only. Use of the AC/DC Adapter charger (Cat. No. 886111) or DC Power cord (Cat. No. 884500) for charging or operation
inside an aircraft must be avoided.
Cat. No. 880051/880061 are not approved for use in aircrafts.
Not intended for use in MRI environments.
Disconnect the LCSU 4 from external power prior to cleaning.

Limited Warranty

The LCSU 4 comes with a two (2) year limited warranty, excluding the Canisters, Tubing systems and Battery. Refer to the
Laerdal Global Warranty for terms and conditions. For more information visit www.laerdal.com. The Battery is warranted for 90
days. Laerdal does not provide Service Parts for this product. Excluding the internal Battery, there are no user-serviceable/user-
replaceable parts inside the LCSU 4 Main Unit.

2 Description and Intended Use

5

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

880061880062880051880052

Table des Matières