Product Specifications - Mettler Toledo InPro 8600i Serie Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

24

5 Product specifications

Specifications InPro 8600 i
© 07 / 16 Mettler-Toledo GmbH, CH-8902 Urdorf
Printed in Switzerland
Turbidity
Measuring principle
Wavelength
Measuring range
Accuracy
Repeatability
Resolution
Response time (T90)
Units
Factory calibration
Colour (InPro 8600 i / D3)
Measuring principle
Wavelength
Measuring range
Measurement accuracy
Repeatability
Resolution
Response time (T90)
Units
Factory calibration
Process conditions
Admissible pressure range
Temperature (process media)
Admissible temperature range (ambient)
Scattered light, red light
InPro 8600 i / D1: 25° (single-angle version)
InPro 8600 i / D3: 25° / 90° (dual-angle version)
650 nm (light-emitting diode [LED])
0 ... 400 FTU
0.01 FTU (measuring value < 1 FTU)
1 % of the measured value (meas. value > 1 FTU)
0.01 FTU
0.01 FTU
< 2 sec
FTU, NTU, EBC, ASBC, ppm SiO
12-point calibration over the entire measuring
range, based on Formazin standards
Absorption, blue light
460 nm (light-emitting diode [LED])
0 – 30 EBC
± 2 % of the measured value
0.01 EBC
0.01 EBC
3 sec at EBC change >10
EBC
6-point calibration
max. 16 bar (232 psi)
InPro 8600 i / D1:
0...100 °C (32...212 °F)
Max. 120 °C (248 °F), short time, 1 h
InPro 8600 i / D3:
0...80 °C (32...176 °F), at 25 °C (77 °F) ambient
temperature with air cooling
0...70 °C (32...158 °F), at 25 °C (77 °F) ambient
temperature without air cooling
Max. 120 °C (248 °F), short time, 1 h
0 ... 60 °C (32 ... 140 °F)
InPro 8600 i Series
, mg/L SiO
2
2
InPro 8600 i Series
30 113 529