6
2

Sensor

2.1

Sensor Certification Specifications

Testing Laboratory,
Types of Protection
and Area Classification
FM Certification
I.S. Class 1 Div 1
Group A – G
T5 at maximum ambient
temperature of 85 ºC
ATEX intrinsically safe for
II 1 GD EEx ia IIC T3...T6
According to EC-type test
certificate KEMA 03 ATEX 1462 X
Table 1: Sensor Certification Specifications
Note: These sensors have been designed to meet the electrical
safety descriptions noted In the table above. For detailed infor-
mation contact your METTLER TOLEDO representative.
2.2

Sensor Identification

A sensor can be identified by the model code located on both
ends of the cable. Refer to «Sensor Applications» section for sen-
sor application information.
Date code
The date code identifies the year and week of manufacture (see
Figure 1). In the example 2B0312 XXX serial 03 identifies the
year of manufacture as 2003, and 12, the week of manufactu-
re in that year. The additional numbers are the serial number.
Entity parameters as follows:
U
i
P
i
The effective internal capacitance C
InPro 7200 Series
52 001 408
Conditions of
Certification
Transmitter CondI7100/2(X)H
Transmitter CondI7100/2(X)H
and M 700X
Figure 1: Sensor Identification
= 7.5 V
= 0.4 W
© 08/04 Mettler-Toledo GmbH, CH-8606 Greifensee
6 m (20 ft)
Agency label
Image of label
I
= 70 mA
i
L
= 7.8 mH
i
is negligibly small.
i
InPro 7200 Series
Suitable Sensor
Types
InPro 7200
InPro 7201
InPro 7202
InPro 7200
InPro 7201
InPro 7202
Model label
Image of label
Printed in Switzerland