11. Electromagnetic compatibility
Guidance and Manufacturer's declaration-electromagnetic emissions
The Luxal is intended for use in the electromagnetic environment specified below . The customer or the user of Luxal should assure
that it is used in the following environment:
Emissions Test
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's declaration-electromagnetic immunity
The Luxal is intended for use in the electromagnetic environment specified below . The customer or the user of the Luxal should
assure that it is used in the following environment:
Immunity Test
IEC 60601-1-2 Test
Level
Electrostatic discharge
±6 kV contact
(ESD)
IEC 61000-4-2
±8 kV air
Electrical fast transient/
±2 kV for power
burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to line(s)
IEC 61000-4-5
±2 kV line(s) to earth
Voltage dips, short
<5% Un
interruptions, and voltage
(>95% drop in Un) for
0,5 cycle
variations on power
supply input lines
40% Un
IEC 61000-4-11
(60% drop in Un) for
5 cycles
70% Un
(30% drop in Un) for
25 cycles
<5% Un
(>95% drop in Un) for
5 seconds
Power frequency (50/60
3 A/m
Hz) magnetic field
IEC 61000-4-8
NOTE Un is the a .c . mains voltage prior to application of the test level
Compliance
Electromagnetic Environment - guidance
The Luxal uses RF energy only for its internal function . Therefore, its RF emissions
Group 1
are very low and are not likely to cause any interference in nearby electronic
equipment .
The Luxal is suitable for use in all establishments, including domestic
Class B
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes .
Class A
Complies
Compliance Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5% Un
(>95% drop in Un) for
0 .5 cycle
40% Un
(60% drop in Un) for
5 cycles
70% Un
(30% drop in Un) for
25 cycles
<5% Un
(>95% drop in Un) for
5 seconds
3 A/m
Luxal Brancard de douche | Mode d'emploi | 59
Electromagnetic Environment - guidance
Floors should be made of wood, concrete, or ceramic
tile . If floors are covered with synthetic material, the
relative humidity should be at least 30% .
Mains power quality should be that of a typical
commercial or hospital environment .
Mains power quality should be that of a typical
commercial or hospital environment .
Mains power quality should be that of a typical
commercial or hospital environment . Luxal is
equipped with a battery .
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment .