Cylinder Fill Times; Specifications - Invacare HomeFill II IOH200AW Manuel D'utilisation

Compresseur
Masquer les pouces Voir aussi pour HomeFill II IOH200AW:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 107

8.2 Cylinder Fill Times

CYLINDER
Concentrator Flow Rate to Patient:
FILLING
Platinum 5 up to 2.5 L/min.
TIMES
Perfecto
up to 2.5 L/min.
2
Platinum 9 up to 5 L/min.
HF2PCE9A
2 hr 20
NORMAL
min
2 hr 45
MAXIMUM
min
All filling times are approximate and may vary on
environmental conditions.
1195663-A
HF2PCE6A
1 hr 25
min
1 hr 50
min

8.3 Specifications

Regulatory Listings:
Electrical Ratings:
Operating Environmental
Conditions:
Atmospheric Pressure:
Storage and Transport:
Input Pressure Required:
Input Flow Required:
Input O
% Required:
2
Width:
Height:
Depth:
Weight:
Shipping Weight:
Max Applied Part Temperature
Technical data
EN/IEC 60601–1
EN/IEC 60601–1–2
230 V~, 1.0 A, 50 Hz
41 to 104° F (5 to 40° C) at
15 to 60% non-condensing
humidity
700 to 1060 hPa
-25 to 158° F (-23 to 70° C)
at 15 to 95% non-condensing
humidity.
14 - 21 psi (96.5 – 144.7 kPa)
2 L/min
>90% O
2
20 ¼ inches (51.44 cm)
15 inches (38.1 cm)
16 inches (40.64 cm)
33 lbs (14.97 kg)
39 lbs (17.69 kg)
122° F (50° C)