Liberator
Table 3
Guidance and manufacturers declaration—electromagnetic immunity
The Liberator is intended for use in the electromagnetic environment specified below. The customer or
the user of the Liberator should assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
IEC 610004-4
input/output lines No data input/output lines
Surge
IEC 61000-4-5
Voltage dips,
<5% UT (>95% dip
short interruptions in UT) for 0,5 cycle
and voltage
40% UT (60% dip
variations on
in UT) for 5 cycles
power supply
70% UT (30% dip
input lines
in UT) for 25 cycles
IEC 61000-4-11
<5% UT (>95% dip
in UT) for 5 sec
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
* This statement indicates that the required testing was performed in a controlled environment and the
Liberator are found to be compliant with regulations.
12 - ENG
PN MN234-C4 B | User Manual
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
Not applicable
supply lines
DC powered device
±1 kV for
Not applicable
±1 kV line(s)
to line(s)
Not Applicable
±2 kV line(s)
DC powered device
to earth
Not Applicable
DC powered device
3 A/m
Electromagnetic environment—guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.*
Not applicable
Not Applicable
Not Applicable
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.