Safety
Caution: Medical Electrical Equipment needs special precautions regarding Electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC infor-
mation provided in this manual.
Caution: Portable and mobile radio frequency (RF) communications equipment can affect
Medical Electrical Equipment.
Caution: The use of Accessories, transducers, and cables other than those specified, with the
exception of transducers and cables sold by the Manufacturer of this device as replacement
parts for internal components, may result in increased Emissions or decreased Immunity of the
Liberator Reservoir.
Caution: The Liberator Reservoir should not be used adjacent to or stacked with other equip-
ment and that if adjacent or stacked use is necessary, the Liberator Reservoir should be
observed to verify normal operation in the configuration in which it will be used.
Table 1
Guidance and Manufacturer's declaration—electromagnetic emissions
The Liberator is intended for use in the electromagnetic environment specified below. The customer or
the user of the Liberator should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions CISPR 11 Class B
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 2*
Recommended separation distances between portable and mobile
The Liberator is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Liberator can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (trans-
mitters) and the Liberator as recommended below, according to the maximum output power of the com-
munications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
(d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 at 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
* This table is included as a standard requirement for equipment which has been tested to specific test
levels and over specific frequency ranges and been found compliant with regulations.
Compliance
Electromagnetic environment—guidance
Group 1
The Liberator uses RF energy only for internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The Liberator is suitable for use in all establishments, including
Not applicable domestic establishments and those directly connected to the
public low-voltage power supply network that supplies
Not applicable buildings used for domestic purposes.
RF communications equipment and the Liberator
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d=1.2√P
0.12 m
0.38 m
1.2 m
3.8 m
12 m
m
80 MHz and 800 MHz 800 MHz to 2,5 GHz
d=1.2 √P
0.12 m
0.38 m
1.2 m
3.8 m
12 m
PN MN234-C4 B | User Manual
Liberator
d=2.3 √P
0.23 m
0.73 m
2.3 m
7.3 m
23 m
11 - ENG