Surface temperature value does not agree with expected
respond. In some cases, the surface being measured is very rough
surface temperature or surface temperature probe is slow to
and a good thermal connection from the probe tip to the surface is
not possible. If possible place a small drop of water on the surface
and then place the probe tip into the water drop. Quick and accurate
results can be obtained in this manner.
Conforms to: ISO 8502-4, BS7079-B and ASTM D3276
Surface
Temperature
Air
Temperature
Humidity
Dew Point
Temperature
*Dependent on RH. See www.defelsko.com/TdAccuracy for more information.
Memory:
Logging Measurement Interval: Adjustable from 1 min – 8 hours
Interface: BLE Bluetooth Smart
IP Rating: 65
Power Source: one CR2032X coin cell
Battery Life^: @ 5 min interval = 120 days
^ The stated battery life is for unattended logging. Battery life is reduced while
connected to the PosiTector App or instrument.
Technical Data
Technical Data
Range
-10° to 80° C
80° to 100° C
14° to 175° F
175° to 212° F
-10° to 60° C
14° to 140° F
0 to 100%
-60° to 60° C
-76° to 140° F
10,000 datasets
@ 5 min interval = 34 days
@ 10 min interval = 68 days
@ 1 hour interval = 200 days
Accuracy
±0.5° C
±1.5° C
±1° F
±3° F
±0.5° C
±1° F
±3 %
±3° C*
±5.4° F*
11
Resolution
0.1° C
0.1° C
0.1° F
0.1° F
0.1° C
0.1° F
0.1 %
0.1° C
0.1° F