3. Reinstall the probe guard by sliding it carefully over the sensor and
then threading it onto the cable/probe assembly with a clockwise
rotation (Figure 2).
4. Locate the gray calibration/storage sleeve that was shipped with
your probe/cable assembly. Moisten the sponge in the gray
calibration/storage with a small amount of clean water.
5. Slide the calibration/storage sleeve over the probe guard to keep
the probe in a moist environment for storage and calibration
(Figure 3). It is important to always keep your sensor in a moist
environment so the sensor cap does not dry out.
6. Connect the cable to the instrument by aligning the connectors,
pushing them together and then screwing down the retainer
clockwise.
Figure 2

BATTERY INSTALLATION

To install (or replace) the battery, remove
the two battery cover screws and battery
cover. Replace the 9V battery. Replace the
battery cover and o-ring (be sure to align
the o-ring correctly to prevent a bad seal)
and fasten the two battery cover screws.
An initial display of "LOW BAT" on the
LCD indicates approximately one hour of
battery life for unit operation within specifications. Replace battery
when "LOW BAT" appears on the LCD.
Battery Disposal
This instrument is powered by a 9 volt battery, which the user must
remove and dispose of when the battery no longer powers the
instrument. Disposal requirements vary by country and region, and
users are expected to understand and follow the battery disposal
requirements for their specific locale.
Figure 3
3
Figure 4